- BSA002 - BetaStar Advanced for Beta-Lactams, Safety Data Sheet [English]
The BetaStar Advanced for Beta-lactams kit contains:
- 100 test strips for detecting beta-lactams stored in tube with desiccant in the lid
- 100 test cartridges
- 100 disposable 400 μL poly-pipettes
- Raptor Integrated Analysis System (Neogen item number: 9680)
- Dymo Label Writer 450 Printer (Neogen item number: 9687)
- Dymo replacement paper (Neogen item number: 9688)
- Raptor Verification Standards (cartridges containing test strip standards) (Neogen item number: 9685)
- BetaStar Advanced Positive Control (Penicillin G at 5.0 ± 0.5 ppb; Desfuroylceftiofur at 100 ± 10 ppb) (Neogen item number: BSA010)
| ppb |
Amoxicillin |
Ampicillin |
Ceftiofur |
Cephapirin |
Cloxacillin |
Penicillin |
| 0 |
0% |
0% |
0% |
0% |
0% |
0% |
| 1 |
|
|
|
|
|
0% |
| 2 |
|
0% |
|
|
0% |
0% |
| 3 |
|
|
|
|
|
0% |
| 4 |
0% |
0% |
|
|
0% |
90% |
| 5 |
|
|
|
|
|
100% |
| 6 |
|
27% |
|
|
0% |
|
| 7 |
27% |
97% |
|
|
|
|
| 8 |
73% |
100% |
|
|
50% |
|
| 10 |
100% |
100% |
|
33% |
100% |
|
| 14 |
|
|
|
73% |
|
|
| 18 |
|
|
|
100% |
|
|
| 20 |
|
|
|
100% |
|
|
| 50 |
|
|
0% |
|
|
|
| 60 |
|
|
13% |
|
|
|
| 80 |
|
|
80% |
|
|
|
| 90 |
|
|
97% |
|
|
|
| 100 |
|
|
100% |
|
|
|
| Tolerance/safe level (ppb) 1 |
10 |
10 |
100 |
20 |
10 |
5 |
| Calculated 90 – 95% 2 |
9.24ppb |
8.57 ppb |
92.66 ppb |
18.07 ppb |
9.03 ppb |
4.63 ppb |
1 The tolerance level for each of the tetracycline antibiotics has been set by the FDA as the sum of all tetracyclines.
2 The 90/95 sensitivity level was calculated from data collected during validation testing by a third party laboratory. Sensitivity was based on testing multiple levels of each tetracycline antibiotic samples per concentration.